Varian—ARIA Radiation Therapy Management Prescribe Treatment Software: Organ at Risk Dose-Volume Constraint Values May Be Displayed Incorrectly If Prescription Was Created in Previous Version

September 21, 2016 | Strategic Insights for Health System

Preview

​ECRI Institute's complete weekly summary of medical device hazard and recall information is available in ECRI Institute's Health Devices Alerts(HDA). For more information about HDA, contact us at clientservices@ecri.org.

In an August 15, 2016, Technical Advisory letter submitted by an ECRI Institute member hospital, Varian states that the above software may incorrectly store and display organ at risk (OAR) dose-volume constraint values if the prescription was created in software versions 11.0, 13.0, or 13.5. Varian also states that this problem occurs because the storage location for dose and volume constraints is swapped in the above software version 13.6 and, as a result, the display of the OAR dose-volume constraints will differ depending on the version in which they were created and the workspace version in which they are displayed. For an example of how the OAR dose-volume constraint values are displayed in software versions 11.0, 13.0, or 13.5 versus software version 13.6, see Table 1 in the letter. Varian further states that this problem does not affect the storage or display of the mean and maximum OAR dose values or Prescription Coverage constraints. Varian states that this problem is unlikely...