GE—Integrated ECG Cables: ​May Short Circuit during Defibrillation and Conduct 25% of Defibrillation Energy Away from the Patient

March 6, 2019 | Strategic Insights for Health System

Preview

​ECRI Institute's complete weekly summary of medical device hazard and recall information is available in ECRI Institute's Health Device Alerts. For more information contact us at clientservices@ecri.org.

In an Urgent Field Safety Notice letter posted by the German Federal Institute for Drugs and Medical Devices (BfArM), GE states that the above cables may short circuit during defibrillation and conduct 25% of the defibrillation energy away from the patient. If this problem occurs during a defibrillation event, it is not noticeable to the caregiver and could contribute to an adverse patient outcome. GE also states that it has received...

Access Full Content