Evaluation: Philips Healthcare MobileDiagnost wDR Mobile Radiographic Unit

December 23, 2015 | Evaluations & Guidance

Preview

  • Primary purpose of the device:
    • General adult 2-D radiographic studies; mostly chest x-ray
    • Pediatric; neonatal
    • Orthopedic
  • Major device components:
    • Motorized wheeled cart with telescopic column
      • The cart is made by Sedecal Inc 374385, and is similar to the cart used by a few other manufacturers.
      • The cart size is 67 × 138 cm.
      • The column is 196 × 15.5 cm.
      • It has decent driving visibility. An object 70 cm high and 20 cm wide can be seen 150 cm from the cart with a small tilt of the user's head.
    • Batteries (1) The system has two separate sets of batteries: one for exposures/computer, and one for the motor drives.

Access Full Content

Contact us today at 610.825.6000.