- Primary purpose of the device:
- General adult 2-D radiographic studies; mostly chest x-ray
- Pediatric; neonatal
- Orthopedic
- Major device components:
- Motorized wheeled cart with telescopic column
- The cart is made by Sedecal Inc 374385, and is similar to the cart used by a few other manufacturers.
- The cart size is 67 × 138 cm.
- The column is 196 × 15.5 cm.
- It has decent driving visibility. An object 70 cm high and 20 cm wide can be seen 150 cm from the cart with a small tilt of the user's head.
- Batteries (1) The system has two separate sets of batteries: one for exposures/computer, and one for the motor drives.