Executive Summary

Robotically assisted surgery has surged in popularity over the last decade and demonstrates significant patient benefits: reduced blood loss, smaller incision, and reduced duration of hospital stay.1-3 However, basic safety steps must still be taken. Although retained foreign objects might be considered less likely during a robotically assisted procedure, the literature demonstrates how discrepancies during robotically assisted surgical procedures can be affiliated with the unintentional retention of a foreign object.4 ECRI Institute PSO has seen cases in which the use of a robot was related to unanticipated variation in perioperative procedures, which may have led to the retention of a foreign object.

Who Should Read This

Table of Contents

PDF version

Robotically assisted surgery has surged in popularity over the last decade and demonstrates significant patient benefits: reduced blood loss, smaller incision, and reduced duration of hospital stay.1-3 However, basic safety steps must still be taken. Although retained foreign objects might be considered less likely during a robotically assisted procedure, the literature demonstrates how discrepancies during robotically assisted surgical procedures can be affiliated with the unintentional retention of a foreign object.4 ECRI Institute PSO has seen cases in which the use of a robot was related to unanticipated variation in perioperative procedures, which may have led to the retention of a foreign object.

Simply put, robots change the dynamic of an otherwise standardized environment in the operating room (OR). Facility and surgical leadership should examine the OR in which the robot is to be used to determine if there is an adequate amount of space for the robot, staff, patient, and all equipment, instruments, and supplies. Unlike laparoscopic or other device-assisted surgeries, robots’ large physical presence can affect staff members’ ability to move around the OR, so this should be examined and addressed. Also, during robotic surgery, it may be necessary to convert to a “traditional” surgical procedure (e.g., open surgery). These procedures require different instruments, and it is important to have the space to make this equipment readily available.1
Perioperative instrument-count safety procedures—namely, instrument counts before, during, and after the procedure and open communication among the surgical team throughout the entire process—must be implemented during robotically assisted surgeries. If uncounted objects are used, they should still be noted and included in review upon the final time-out. Typically, the unintentional retention of foreign objects is more likely to occur under the following circumstances: when the surgery is emergent, when there are multiple procedures, when an unexpected change must occur during the procedure, or when the patient has a high body mass index.5 Perioperative communication breakdown and inconsistent adherence to evidence-based practices compound the issue. Regardless of the type of procedure being performed, surgical staff leadership should review and consistently enforce perioperative strategies to prevent retention of foreign objects.

Key Contributing Factors

  • Management/organization: inconsistent enforcement of perioperative safety practices5
  • Team coordination: unclear or nonexistent communication4
  • Operating environment: logistical considerations when using a robot1
  • Workflow: disempowerment of surgical team members6

Key Recommendations

  1. Before beginning a robotically assisted surgery program, include a simulated procedure with the robot in the environment where it will be used as part of staff training to attain a better understanding of how its presence will affect the layout and workflow of the OR. Third-party observation and postprocedure team debriefing should also be considered. Alternatively, perform a detailed assessment of how the robot will change the workflow.
  2. Consider that supportive techniques, such as imaging or radiofrequency tagging, can help maintain an accurate perioperative inventory. In most cases of retained foreign objects in the literature, counts seemed correct even though an object was found in the patient.6-8
  3. Review the organization’s culture, and empower all staff members to stop a procedure if a discrepancy in instrument counts or a missing component of a surgical device is noted.6 Ensure that communication among staff during procedures is open and clear

Take Home Point

When preparing for or performing robotically assisted surgery, the organization needs to assess how a robot’s physical presence will alter staff movement, workflow, and communication during a surgical procedure. An environmental assessment can help determine optimal human interactions with robots.

Director's Note

The surgery team’s dynamic and individual surgical techniques may change with the use of robots and introduction of other technologies, but steps to assess the environment and maintain perioperative safety practices are key to ensuring patient safety. A culture of safety—a concept ECRI Institute has always supported—is critical in the side-by-side evolution of practice and safety. If your institution has experienced a case in which a foreign object was unintentionally retained in a patient, we can provide confidential assistance to determine why this occurred. Patient Safety E-lerts provide participating organizations with additional periodic educational awareness to help prevent healthcare events from happening in their facilities. To discuss your OR safety concerns, please contact us at (610) 825-6000 or patientsafety@ecri.org, and we will forward your questions to our experts.

Karen P. Zimmer, M.D., M.P.H., FAAP, Clinical Director

Glossary

Bibliography

References

  1. Holloway RW, Patel SD, Ahmad S. Robotic surgery in gynecology. Scand J Surg 2009;98(2):96-109.
  2. Maggioni A, Minig L, Zanagnolo V, et al. Robotic approach for cervical cancer: comparison with laparotomy: a case control study. Gynecol Oncol 2009 Oct;115(1):60-4.
  3. Veljovich DS, Paley PJ, Drescher CW, et al. Robotic surgery in gynecologic oncology: program initiation and outcomes after the first year with comparison with laparotomy for endometrial cancer staging. Am J Obstet Gynecol 2008 Jun;198(6):679:e1-10.
  4. Sakhel H, Hines J. To forget is human: the case of the retained bulb. J Robotic Surg 2009 Mar;3(1):45-7.
  5. ECRI Institute PSO. Prevention of retained foreign objects. PSO Navigator 2010 Aug;1(2):1-3.
  6. Beyond the count: preventing the retention of foreign objects. Pa Patient Saf Advis [online] 2009 Jun [cited 2012 Mar 20]. Available from Internet: http://patientsafetyauthority.org/ADVISORIES/AdvisoryLibrary/2009/Jun6(2)/Pages/39.aspx.
  7. American College of Surgeons. Statement on the prevention of retained foreign bodies after surgery. 2005 [cited 2012 Mar 20]. Available from Internet: http://www.facs.org/fellows_info/statements/st-51.html.
  8. NoThing Left Behind: a national surgical patient-safety project to prevent retained surgical items [website]. [cited 2012 Mar 20]. San Francisco: Gibbs VC. Available from Internet: http://www.nothingleftbehind.org.

Resource List

Related Resources

Topics and Metadata

Topics

Quality Assurance/Risk Management

Caresetting

Hospital Inpatient; Hospital Outpatient; Ambulatory Surgery Center

Clinical Specialty

Surgery

Roles

Clinical Practitioner; Nurse

Information Type

Alerts

Phase of Diffusion

 

Technology Class

 

Clinical Category

 

UMDNS

SourceBase Supplier

Product Catalog

MeSH

ICD9/ICD10

FDA SPN

SNOMED

HCPCS

Disease/Condition

 

Publication History

​Published May 31, 2012